Health Products

Caps for Capillary (2015-01-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots made after April 2013.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. Caps for Capillary10311053

Introduction

Recalled products
  1. Caps for Capillary

Reason

Following receipt of customer complaints, an investigation identified that a material change made by the supplier has resulted in capillary caps that are more difficult to apply to the capillaries. Attempting to apply these capillary caps to the capillaries could result in spillage of the sample or possible breakage of the glass capillary. Siemens is advising customers to use caution when applying the capillary caps with Siemens multicap glass and multicap-s plastic capillaries to minimize the potential for breakage or bending.