ADVIA Centaur and Centaur CP Systems (2015-01-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
42564268
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. ADVIA Centaur System - Progesterone Assay01586287
42564268
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
A. ADVIA Centaur CP System - Progesterone (PRGE) Assay01586287

Introduction

Recalled products
  1. ADVIA Centaur CP System - Progesterone (PRGE) Assay
  2. ADVIA Centaur System - Progesterone Assay

Reason

There is a high bias on results greater than 30 ng/ml (95.4 nmol/l) when compared to the master curve when using select ADVIA centaur progesterone kit lots. The high bias may cause neat serum samples with results greater than 60 ng/ml (190.8 nmol/l) to recover within the assay range upon dilution (dilution factor applied). Samples with values below 30 ng/ml (95.4 nmol/l) do not demonstrate this bias.