Indiko and Indiko Plus (2014-02-28) Feb 28, 2014 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number Not applicable Manufacturer Thermo Fisher Scientific Oy P.O. Box 100, Ratastie 2 Vantaa 01621 FINLAND B) Indiko Plus 9864 0000-0 Not applicable Manufacturer Thermo Fisher Scientific Oy P.O. Box 100, Ratastie 2 Vantaa 01621 FINLAND A) Indiko 98630000-0 INDIKO
Recalled Products
A) Indiko
B) Indiko Plus
A new cuvette does not always replace the old used cuvette from the incubator. As a result, a new sample might be dispensed to an already used cuvette causing an erroneous result.