CLINITEK Novus Automated Urine Chemistry Analyzers (2015-01-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Siemens Healthcare Diagnostics Manufacturing Ltd.,
Northern Road,
Sudbury, Suffolk, United Kingdom
CO10 2XQ
B) CLINITEK Novus Automated Urine Chemistry Analyzer (Without Rack Handler)10697937
All lots.
Manufacturer
Siemens Healthcare Diagnostics Manufacturing Ltd.,
Northern Road,
Sudbury, Suffolk, United Kingdom
CO10 2XQ
A) CLINITEK Novus Automated Urine Chemistry Analyzer (With Rack Handler)10494134

Introduction

Recalled Products

A) CLINITEK Novus Automated Urine Chemistry Analyzer (With Rack Handler)

B) CLINITEK Novus Automated Urine Chemistry Analyzer (Without Rack Handler)

Reason

Due to a transcription error in the CLINITEK Novus software, there is a potential for protein to creatinine (P:C) ratios of > 300 mg/g to be reported higher than expected. In particular:
  1. a p:c ratio of 300 mg/g may be reported as 500 mg/g.
  2. a p:c ratio of 500 mg/g may be reported as >1500 mg/g.


This error can potentially occur when a CLINITEK Novus PRO12 urinalysis cassette is used on the CLINITEK Novus Analyzer.
P:C ratios reported as normal or 300 mg/g are not affected.