Health Products

Allura XPER FD10 (2015-01-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A) Allura XPER FD10ALLURA XPER FD10

Introduction

Recalled Products

A) Allura XPER FD10

Reason

Philips has received a customer feedback where the monitor ceiling suspension system fell to its lowest position. In this specific occasion it collided with the table top. The investigation initiated concluded that the cause of the failure was an assembly error of the actuator of the MCS.