Zenith Alpha Thoracic Endovascular Graft (proximal, proximal tapered, and distal components) (2015-01-30) Jan 30, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number All lots Manufacturer William Cook Europe APS Sandet 6 Bjaeverskov 4632 DENMARK B. Z.Alpha thor. endo. graft proximal tapered More than 10 numbers, contact manufacturer. All lots Manufacturer William Cook Europe APS Sandet 6 Bjaeverskov 4632 DENMARK C. Zenith Alpha thor.endo.graft distal component More than 10 numbers, contact manufacturer. All lots Manufacturer William Cook Europe APS Sandet 6 Bjaeverskov 4632 DENMARK A. Z.Alpha T. Endo Graft proximal component More than 10 numbers, contact manufacturer.
Recalled products
Z.Alpha T. Endo Graft proximal component
Z.Alpha thor. endo. graft proximal tapered
Zenith Alpha thor. endo. graft distal component
Field safety notice advising users of updates to the IFU that are being made to emphasize best practices in response to a few recent findings of distal Type I endoleak, migration and aneurysm growth during on-going longer term follow up of patients enrolled in the multi national clinical investigation of the device.