Health Products

LABType HD DRB1 Typing Test

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
009, BATCH 1-16A. LABType HD DRB1 Typing Test
Manufacturer
One Lambda Inc
21001 Kittridge Street
Canoga Park
California
UNITED STATES
RSSOH2B1

Introduction

Recalled products
  1. LABType HD DRB1 Typing Test

Reason

The following allele specificities are missing from the lot specific user documents and catalog file for the LABType HD DRB1 typing test (Catalog ID RSSOH2B1, LOT 009, Batches 1-16):
  • DRB1*03:12
  • DRB1*03:16
  • DRB1*04:23
  • DRB1*04:25
  • DRB1*04:32
  • DRB1*04:52
  • DRB1*11:40
  • DRB1*14:142
  • DRB1*15:06:02
  • DRB1*15:74
  • DRB1*15:85
  • DRB3*03:02
  • DRB4*01:05
  • DRB4*01:08


This may cause a mistype for DRB1*04:52.  A sample with DRB1*04:52 would be typed as DRB1*04:03:01/03:03/03:05/03:07~03:08 without indicating an ambiguity with DRB1*04:52.