Health Products

ALLSET Gold DRB Class II High Resolution Kits and SSP Unitray (2015-02-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
015 1322626
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
Wisconsin
UNITED STATES
B. DRB1-13 SSP UNITRAY 12 TESTS45080-4
015 1446996
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
Wisconsin
UNITED STATES
A. ALLSET GOLD DRB1-13 HIGH RES54270D

Introduction

Recalled Products

A. ALLSET GOLD DRB1-13 HIGH RES
B. DRB1-13 SSP UNITRAY 12 TESTS

Reason

The Primer mix in Lane 6 (Primer Mix R13-51B) of the affected products may cause a DRB1*13:50 allele to be typed as a DRB1*13:14 allele. These alleles occur very rarely. This issue is due to limited sequence information available for this allele at the time of release. Due to the recent availability of a sample exhibiting the DRB1*13:50 allele, a new lot-specific comment has been added to the quality control comments alerting users to the possibility of a false negative in Lane 6 with the DRB1*13:50 allele.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime