IMMULITE 2000 System - GI-MA ASSAY (2015-02-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
LOTS 312 , 313IMMULITE 2000 System - GI-MA ASSAY
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
L2KGI2

Introduction

Recalled Products

IMMULITE 2000 System - GI-MA ASSAY

Reason

Siemens Healthcare Diagnostics has confirmed a positive bias with the Bio-Rad Lyphochek tumor marker plus control and the Bio-Rad Liquichek tumor marker control when used with Immulite 2000/Immulite 2000 XPI GI-MA (CA19-9) assay kit Lots 312 and 313. In addition, a greater than expected change in patient sample recovery may be observed when moving from kit Lot 311 to kit Lots 312 and above. While recovery with patient samples for individual kit Lots continues to meet Siemens quality control release specification, kit Lot 311 was observed to perform at the lower end of the specification and kit Lot 312 at the upper end of the specification, resulting in a greater than expected difference between these two kit Lots.