Health Products

AQUARIUS Workstation Basic Unit (2015-01-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
AQ-UH4940
Manufacturer
TeraRecon, Inc.
4000 East 3rd Avenue
AQUARIUS Workstation Basic Unit84-0069

Introduction

Recalled Products

AQUARIUS Workstation Basic Unit

Reason

Software anomaly related to the Recist 1.1 target lesion evaluation criteria in the findings workflow module within the aquarius workstation intuition client viewer. In certain instances, this target lesion evaluation output provided by the software may be inaccurate. The total measurement calculation and percentage of change are accurate. The affected versions are 4.4.11.82.6784, 4.4.11.116.7134 and 4.4.11.144.7589.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime