Health Products

LIASON XL Analyzer and XL System Analyzer (2015-02-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2210000392
Manufacturer
Diasorin Inc.
1951 Northwestern Avenue
P.o. Box 285
Minnesota
UNITED STATES
B. LIAISON XL System - AnalyzerI0050
2210000392
Manufacturer
Diasorin Inc.
1951 Northwestern Avenue
P.o. Box 285
Minnesota
UNITED STATES
A. LIAISON XL AnalyzerI0050

Introduction

Recalled Products

A. LIAISON XL Analyzer
B. LIAISON XL System - Analyzer

Reason

A correction to parameter #61 of the Washer Dispense Pump (WDP) on the Liaison XL Analyzer is required to prevent unnecessary Washer Aspiration Failures (WAF) and Target Not Reached errors due to low dispense volume.