Health Products

OXOID LEGIONELLA BCYE GROWTH SUPPLEMENT (2015-01-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1545171
Manufacturer
Oxoid Limited
Wade Road
Basingstoke
OXOID LEGIONELLA BCYE GROWTH SUPPLEMENTSR0110A

Introduction

Recalled Products  

OXOID LEGIONELLA BCYE GROWTH SUPPLEMENT

Reason

A technical investigation has found that specific Lots of Oxoid Legionella BCYE growth supplement (SR0110A) may not deliver adequate recovery of Legionella Pneumophila when used in conjunction with Oxoid Legionella medium (CM0655) and also with Oxoid Legionella selective supplement variants. Continued use of these Lots could result in false negative reporting.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime