175913B. Diluent Universal
Manufacturer
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
03183971122
179689A. CMV IGM for ELECSYS and COBAS E
Manufacturer
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
04784618190

Introduction

Recalled Products

A. CMV IGM for ELECSYS and COBAS E
B. Diluent Universal

Reason

This recall is related to a potential reduction in the signal of the Elecsys CMV IGM assay when used in combination with some specific Lots of Diluent Universal which may lead to false negative results for patients samples with high reactivity against one specific antigen. This may be affected by reduced cut-off values which may lead to a false negative result when the affected combination is used. The signal reduction has only minor effects on the calibrators and controls.