VERTEX EPIC CAMERA/PEGASUS ULTRA IMAGING SYSTEM (2015-03-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.VERTEX EPIC CAMERA/PEGASUS ULTRA IMAGING SYSTEM
Manufacturer
Philips Medical Systems (Cleveland) Inc.
3860 North First Street
San Jose
95134
UNITED STATES
AMT-VERTEXA

Introduction

Recalled Products 

VERTEX EPIC CAMERA/PEGASUS ULTRA IMAGING SYSTEM

Reason

Philips Healthcare received a report from the field that during a manual collimator exchange procedure, the operator was attempting to slide the collimator cassette containing the detector 2 vertex high resolution (VXHR) collimator from the collimator storage cabinet onto the collimator exchange carriage.  As the VXHR collimator and collimator cassette was being slid onto the carriage, the cassette did not align with the carriage top guide rollers and mechanical lock. The collimator and collimator cassette fell off of the carriage and onto the floor.