Recalled Products
5392 DUAL CHAMBER EXTERNAL PACEMAKER
Medtronic Model 5392 EPGS could experience a performance issue when used with specific AA-sized (LR6) batteries. Medtronic has received six (6) reports (out of approximately 6,000 model 5392 EPGS distributed globally) where the negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This issue could prevent the EPG from powering on or cause the EPG to abruptly lose primary battery power, potentially stopping delivery of pacing therapy. There have been no reports of patient harm as a result of any occurrence of this issue.