Health Products

5392 DUAL CHAMBER EXTERNAL PACEMAKER (2015-03-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
Minnesota
UNITED STATES
5392 DUAL CHAMBER EXTERNAL PACEMAKER5392

Introduction

Recalled Products   

5392 DUAL CHAMBER EXTERNAL PACEMAKER

Reason

Medtronic Model 5392 EPGS could experience a performance issue when used with specific AA-sized (LR6) batteries. Medtronic has received six (6) reports (out of approximately 6,000 model 5392 EPGS distributed globally) where the negative terminal of a commercially available AA (LR6) battery did not maintain a sufficient connection with the battery drawer electrical contact. This issue could prevent the EPG from powering on or cause the EPG to abruptly lose primary battery power, potentially stopping delivery of pacing therapy. There have been no reports of patient harm as a result of any occurrence of this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime