Health Products

T2100 TREADMILL (2015-03-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
GE Medical Systems Information Technologies
Inc.
8200 West Tower Avenue
Wisconsin
UNITED STATES
A. T2100 TREADMILL2020835-001

Introduction

Recalled Products

A. T2100 TREADMILL

Reason

Performance issue with the T2100 Treadmills can cause uncontrolled walking belt motion during a stress exercise test. This motion can be seen as an unexpected slowing, followed by sudden belt acceleration in either the forward or backward direction. When this occurs, engaging the emergency stop button (ESTOP) will not immediately stop the belt. Rather, it will coast to a stop within 35 seconds when slowing from maximum speed. The sudden change in belt speed and/or direction can result in a fall and injury to the patient.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime