B. VITROS 5,1 FS CHEMISTRY SYSTEM - ANALYZER (REFURBISHED)
Manufacturer
Ortho-clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
UNITED STATES
6801890
More than 10 numbers, contact manufacturer.
C. VITROS 5,1 FS CHEMISTRY SYSTEM
Manufacturer
Ortho-clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
UNITED STATES
6801375
More than 10 numbers, contact manufacturer.
A. VITROS 5,1 FS CHEMISTRY SYSTEM - ANALYZER
Manufacturer
Ortho-clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
UNITED STATES
6801375
Introduction
Recalled Products
A. VITROS 5,1 FS CHEMISTRY SYSTEM - ANALYZER
B. VITROS 5,1 FS CHEMISTRY SYSTEM - ANALYZER (REFURBISHED)
C. VITROS 5,1 FS CHEMISTRY SYSTEM
D. VITROS 5,1 FS CHEMISTRY SYSTEM (REFURBISHED)
Reason
Ortho Clinical Diagnostics (OCD) has identified an anomaly with Vitros System software version 3.1 and below. OCD determined that the software may not properly identify an expired calibration. Calibration status is monitored by the system software which is based upon the calibration interval and the date of the last calibration. OCD's investigation confirmed that if the anomaly occurs, a calibration may expire, but the system may not alert the operator and not flag results with a CE sample result code (i.e. calibration expired). The system may unexpectedly generate results even though it is configured to not report results when a calibration has expired.