Health Products

LCS COMPLETE KNEE SYSTEM (2015-03-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
DePuy Orthopaedics Inc.
700 Orthopaedics Drive, P.O. Box 988
Warsaw
46581-0988
Indiana
UNITED STATES
B. LCS COMPLETE KNEE - RPS FLEXION FEMORAL129411010
All lots.
Manufacturer
DePuy Orthopaedics Inc.
700 Orthopaedics Drive, P.O. Box 988
Warsaw
46581-0988
Indiana
UNITED STATES
A. LCS COMPLETE KNEE SYSTEM - RPS INSERTMore than 10 numbers, contact manufacturer.

Introduction

Recalled Products

A. LCS COMPLETE KNEE SYSTEM - RPS INSERT

B. LCS COMPLETE KNEE - RPS FLEXION FEMORAL

Reason

Depuy Orthopaedics Inc., is issuing a device correction for all lots of the LCS complete RPS Knee System and advising surgeons that the product labeling will be revised to include a warning that states the patella must be resurfaced when using this system. According to data from the Australian Orthopaedic Association National Joint Replacement Registry, failure to resurface the patella has been associated with a higher incidence of postoperative patello-femoral pain, potentially leading to a secondary procedure.