Vertex Epic Camera / Pegasys Ultra Imaging System (2015-03-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems (Cleveland) Inc.
3860 NORTH FIRST STREET,
SAN JOSE, CALIFORNIA
95134
UNITED STATES
A. Vertex Epic Camera / Pegasys Ultra Imaging SystemAMT-VERTEXA

Introduction

Recalled Products

A. Vertex Epic Camera / Pegasys Ultra Imaging System

Reason

When the Gamma camera detector #1 head is at a position directly above the patient table (relative 180 degree position), oil and/or grease contamination on the radius drive clutch (ES) may result in non-user initiated drifting down of the gamma camera, detector #1 head towards the patient table. Patient injury may result if, a patient is on the table and the downward drift is not detected in time to execute emergency patient removal.