A. Trellis 6 & Trellis 8 Peripheral Infusion System
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
Massachusetts
UNITED STATES
BVT608010
Introduction
Recalled products
Trellis 6 & Trellis 8 Peripheral Infusion System
Reason
Covidien is issuing a voluntary recall of all Trellis 6 and Trellis 8 peripheral infusion systems due to potential for a sterility breach of the outer packaging or pouch material. A breach of the outer pouch may compromise the outside surface sterility of the inner pouch and does not directly affect the sterility of the device components within the inner pouch.