Health Products

Cobas IT Middleware Core Unit (2015-04-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All software versions.
Manufacturer
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
A. Cobas IT Middleware Core Unit06498027001

Introduction

Recalled products
  1. Cobas IT Middleware Core Unit

Reason

Cobas IT Middleware feature "sample closing" could fail if a "rebuilt sample" message was sent by the host before the "sample closing date" already established by the user. The sample result mismatch may occur, if the customer allows recycling patient's ids and if their system was set to allow "sample rebuild".
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime