Health Products

YSIO MAX (2015-04-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
SN 24102
Manufacturer
Siemens Ag
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A. YSIO MAX10762470

Introduction

Recalled Products

A. YSIO MAX

Reason

Images might be lost and corresponding acquisitions have to be repeated under the following circumstances: 
-during an image recovery process on systems with portable detectors (Max WI-D, Max Mini) the connection of the portable detector is sporadically not re-established.
-if either the "undo" button on the rad subtask card or the "previous series" button on the image subtask card are pressed during image readout.
-sporadically, during an automatic or a manual RIS update. This may cause the deletion of the currently used study and all acquired images within this study.