Intellivue Information centre IX - software (2015-04-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
1M4H-34MB-4A. Intellivue Information centre IX - software
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
Massachusetts
UNITED STATES
M3290B

Introduction

Recalled Products

A. Intellivue Information centre IX - software

Reason

Philips has become aware that when HL7 lab messages are sent to a PIIC iX system that has been localized for comma delimiters, the lab result is incorrectly displayed on the bedside monitor. An expected delimiter is stripped from the lab value and the number shown is of a higher magnitude than expected. If lab results are incorrect, a clinician may initiate incorrect treatment or delay necessary treatment this problem only happens when the PIIC iX is configured during set up for regional settings where the delimiter is a comma. The lab option must be in use, and the workflow utilizes the automatic lab interface displayed at the bedside monitor.