Health Products

Digital Diagnost (2015-04-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
63902199
Manufacturer
Philips Medical Systems Dmc Gmbh
Rontgenstrabe 24
Hamburg
22335
GERMANY
A. Digital Diagnost9890 010 83904

Introduction

Recalled Products

A. Digital Diagnost

Reason

The system is designed to emit an acoustic signal (beep) upon termination of an exposure. Philips has become aware that if the system has been powered on for more than 12 hours, the system will no longer emit this signal. This signal is a requirement as per IEC 60601-1-3, clause 6.4.2 and the radiation emitting devices regulations C.R.C., CH. 1370, part xii, section 6 (g).