Health Products

Softwares for VITEK Systems (2015-02-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Version 7.01B. Software for VITEK 2 Compact 60 System
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
Missouri
UNITED STATES
27560
Version 7.01C. Software for VITEK 2 System- VITEK 2
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
Missouri
UNITED STATES
27225
Version 7.01A. Software for VITEK 2 Compact 30 System
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
Missouri
UNITED STATES
27530

Introduction

Recalled Products
  1. Software for VITEK 2 Compact 30 System
  2. Software for VITEK 2 Compact 60 System
  3. Software for VITEK 2 System- VITEK 2

Reason

VITEK 2 7.01 software introduced a feature (enable - convention for highest reportable MIC) that can be used to modify the reporting convention when an AST result is within the highest reportable mic range.  Advanced expert system (AES) parameter sets are configurable in the VITEK 2 software to allow users to select mic breakpoint interpretation as well as AES functionality related to recommended guidelines. When "enabled - convention for highest reportable MIC configuration option modifies the highest reportable MIC value for each specific antimicrobial to half its doubling dilution and displays the value with a "greater than" symbol.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime