Beckman Coulter has received a number of complaints concerning failed calibrations for specific lots of AU Bicarbonate Reagent used on Beckman Coulter AU analyzers. If calibration failures are generated using the specified lots of bicarbonate reagent, this may cause a delay in generation and release of patient results. No erroneous results will be generated as a result of this failure.