A. Dimension Vista System - Thyroid Stimulating Hormone (TSH) Assay
K6412
Introduction
Recalled Products
A. Dimension Vista System - Thyroid Stimulating Hormone (TSH) Assay
Reason
Siemens Healthcare Diagnostics has determined that Dimension Vista loci Thyroid Stimulating Hormone (TSH) lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a flex well. The occurrence of this issue is low. Quality control will detect the issue if it is run near the end of an affected well set.