Dimension Vista System - Thyroid Stimulating Hormone (TSH) Assay (2015-04-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
14237AA
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 Gbc Drive
Mailstop 514
Delaware
UNITED STATES
A. Dimension Vista System - Thyroid Stimulating Hormone (TSH) AssayK6412

Introduction

Recalled Products

A. Dimension Vista System - Thyroid Stimulating Hormone (TSH) Assay

Reason

Siemens Healthcare Diagnostics has determined that Dimension Vista loci Thyroid Stimulating Hormone (TSH) lot 14237AA may produce falsely depressed results ranging from -23% to -96% near the end of the well set due to inadequate reagent volume in a flex well. The occurrence of this issue is low. Quality control will detect the issue if it is run near the end of an affected well set.