RAPIDlab 1265 System - Analyzer (2015-05-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
14532
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
New York
UNITED STATES
A. RAPIDlab 1265 System - AnalyzerNot applicable

Introduction

Recalled Products
  1. RAPIDlab 1265 System - Analyzer

Reason

During some internal testing at Siemens, it was discovered that the D50 (glucose sensor) and D51 (lactate sensor) error codes were not functioning on the RAPIDLab 1260 and RAPIDLab 1265 instruments. Glucose and lactate sensors that are not completely seated with the contacts on the instrument should alert the operator with a D50 (glucose sensor) error or a D51 (lactate sensor) error.