Some of the Corometrics 171, 172, 173 and 174 series monitors were delivered with the FHR lower alarm limit configured incorrectly at 50 BPM (proper factory default setting is 120 BPM). A lower than expected FHR, if left undetected, could result in fetal bradycardia and oxygen deprivation to the fetus. There have been no injuries reported as a result of this issue.