Health Products

M/DN Intramedullary 4.5MM Diameter Screws (2015-05-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.A. M/DN Intramedullary 4.5MM Diameter  Screws
Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
More than 10 numbers, contact manufacturer.

Introduction

Recalled Products
  1. M/DN Intramedullary 4.5MM Diameter  Screws

Reason

Zimmer is initiating a recall of M/DN Intramedullary Fixation 4.5mm Diameter Cortical Screws listed as COMP17IBLE with the M/DN system as it was found through review of the design history files that they are not compatible with the M/DN system. The compatibility issue can only occur at worst case print specification wherein multiple features are at maximum material condition (meaning the hole in the nail would need to be as small as possible and the outside diameter of the screw would need to be as large as possible).