Oasis Drain, Dry Single W/AC (2015-05-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
L/N 222115
Manufacturer
Atrium Medical Corporation
5 Wentworth Drive
Hudson
New Hampshire
UNITED STATES
A. Oasis Drain, Dry Single W/AC3600-100

Introduction

Recalled Products
  1. Oasis Drain, Dry Single W/AC

Reason

In rare cases the O-ring on the in-line connector may be missing on the end of the patient line. The O-ring is a black circular ring that is clearly visible when the two pieces of the in-line connector are separated.  If this O-ring is missing, then air can pass into the patient tubing causing a system air leak.  The chest drain system will no longer be a closed system as required and you will not obtain appropriate function of the drain in evacuating air, blood or fluid.