Microplate Reader Infinite (2015-06-15) Jun 15, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer. Manufacturer Tecan Austria GmbH, Untersbergstrasse 1A, A-5082, Grodig 9999 AUSTRIA D. Microplate Reader Infinite M1000 30034301 More than 10 numbers, contact manufacturer. Manufacturer Tecan Austria GmbH, Untersbergstrasse 1A, A-5082, Grodig 9999 AUSTRIA E. Microplate Reader Infinite 200 30016056 More than 10 numbers, contact manufacturer. Manufacturer Tecan Austria GmbH, Untersbergstrasse 1A, A-5082, Grodig 9999 AUSTRIA B. Microplate Reader Infinite M 1000 PRO 30063849 More than 10 numbers, contact manufacturer. Manufacturer Tecan Austria GmbH, Untersbergstrasse 1A, A-5082, Grodig 9999 AUSTRIA C. Microplate Reader Infinite F500 30019337 More than 10 numbers, contact manufacturer. Manufacturer Tecan Austria GmbH, Untersbergstrasse 1A, A-5082, Grodig 9999 AUSTRIA A. Microplate Reader Infinite 200 PRO 30050303
Recalled Products
Microplate Reader Infinite 200 PRO
Microplate Reader Infinite M 1000 PRO
Microplate Reader Infinite F500
Microplate Reader Infinite M1000
Microplate Reader Infinite 200
During an internal project Tecan has determined that due to a firmware bug incorrect luminescence measurement values are displayed if the actual luminescence measurement value equals 4096 RLU or multiples thereof. It was determined that these incorrect measurement values may potentially impact the application result.