Health Products

Dako Beta-2-Microglobulin Kit (2015-05-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
DAKO Denmark A/S
Produktionsvej 42
Glostrup, Copenhagen
2600
DENMARK
A. Dako Beta-2-Microglobulin KitOA368

Introduction

Recalled Products

A. Dako Beta-2-Microglobulin Kit

Reason

Dako has requested that Bec distribute the Dako recall notification letter, which has been issued due to the stability of the single component of the kit (Vial 1 Beta-2- microglobulin immunoparticles). According to Dako's notification, investigations have shown that a change in the Ph value in the beta-2-microglobulin immunoparticles has occurred, resulting in low or no signal for the two lowest standards in the calibration curve. In addition, the identified change in the Ph value may affect the accuracy of the beta-2-microglobulin values and results obtained, without any warnings from the instruments.