ENGEN LABORATORY AUTOMATION SYSTEM (2015-05-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
J16XQH4J
Manufacturer
Thermo Fisher Scientific Oy
P.O. Box 100, Ratastie 2
Vantaa
01621
FINLAND
A. ENGEN LABORATORY AUTOMATION SYSTEMENGEN

Introduction

Recalled Products

A. ENGEN LABORATORY AUTOMATION SYSTEM

Reason

Under certain conditions, when an enGen Laboratory Automation System (enGen system) is configured with a recapper module, there is a potential for recapper caps for use with the recapper module to fall off of or be knocked from sample tubes placed in the storage racks of entry/exit modules. There is also the possibility for the gripper to grab a tube at the wrong height and drop the tube in the rack entry/exit or rack exit module. These situations may result in splashing of bio-hazardous fluid, and sample cross-contamination may occur if there is an uncapped tube in the rack.