Health Products

Standard Offset Cup Impactor (2015-05-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Greatbatch Medical Orthopaedics
10000 Wehrle Drive
Clarence
14031
New York
UNITED STATES
A. Standard Offset Cup ImpactorT6318

Introduction

Recalled Products

A. Standard Offset Cup Impactor

Reason

Greatbatch had originally identified through internal sterility assurance level validation testing that a sterility assurance level of 10-6 is not achieved on the standard offset cup impactor when it is subjected to the steam sterilization cycles and associated dry times that are specified in the current Instructions for Use (IFU) provided with the device. The standard offset cup impactor is provided non-sterile and must be sterilized prior to use in a surgery. Stryker orthopedics then took additional risk mitigation steps to include the removal of the blue plastic handle. All modified units (T6318) are now being removed from the field.