Health Products

Karl Strorz Morcellators ( 2015-05-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All serial numbers.
Manufacturer
Karl Storz Gmbh & Co. Kg
Mittelstrasse 8
Tuttlingen
78532
GERMANY
B. RotoCut Morcellator20701020-1
All serial numbers.
Manufacturer
Karl Storz Gmbh & Co. Kg
Mittelstrasse 8
Tuttlingen
78532
GERMANY
A. SAWALHE Morcellator20711120-1

Introduction

Recalled Products

A. SAWALHE Morcellator
B. RotoCut Morcellator

Reason

When a laparoscopic power morcellator is used on a patient with (undiagnosed) cancer, there is a possibility that the cancer may be upstaged, and therefore, shorten the patient's life span. Health Canada posted a safety information notice to hospitals on May 7, 2014 regarding the use of the device and risk of spreading unsuspected uterine sarcoma. Health Canada requested a labelling change under section 39 of the medical devices on December 8, 2014 and posted an updated communication on their website on December 10, 2014. Karl Storz is revising their user manuals in compliance with Health Canada's recommendations.