Dimension Vista System - Pseudocholinesterase Assay (2015-05-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
14147BC                     
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 Gbc Drive, Mailstop 514
Newark, Delaware
19714-6101
UNITED STATES
A. Dimension Vista System - Pseudocholinesterase AssayK3051

Introduction

Recalled Products

A. Dimension Vista System - Pseudocholinesterase Assay

Reason

Increase in frequency of absorbance errors [E141: IOD error] on patient samples with the dimension vista PCHE Flex reagent lots: 14147bc, 14160ba, 14168ab and 14198ac. The [E141: IOD error] flag is generated from an IOD (initial optical density) test value that is less than the software coded limit of 575. The IOD test calculation is used to monitor reagent delivery. Reagent that does not produce the typical colour may also generate lower than normal IOD test values, triggering the [E141: IOD error] flag. The issue does not affect every flex and not all well sets. Only patient samples with values close to the upper end of the assay range will be impacted and the absorbance error [E141: IOD error] flag are not reportable.