ADVIA Centaur System (2015-05-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
New York
UNITED STATES
B. ADVIA Centaur XP system-Troponin I Ultra (TnI-Ultra) Assay02789602
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
New York
UNITED STATES
C. ADVIA Centaur CP system-Troponin I Ultra (TnI-Ultra) Assay2789602         
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
New York
UNITED STATES
A. ADVIA Centaur System-Troponin I Ultra (TnI-Ultra) Assay2789602              

Introduction

Recalled Products

A. ADVIA Centaur system-Troponin I Ultra (TnI-Ultra) Assay
B. ADVIA Centaur XP system-Troponin I Ultra (TnI-Ultra) Assay
C. ADVIA Centaur CP system-Troponin I Ultra (TnI-Ultra) Assay

Reason

Siemens healthcare diagnostics has observed a system-to-system bias between the TnI-Ultrat assay on the ADVIA Centaur CP and ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT systems with the ADVIA Centaur CP generating lower results than the ADVIA Centaur/ ADVIA Centaur XP / ADVIA Centaur XPT system.  This observation impacts customers who use the TnI-Ultrat assay on both the ADVIA centaur CP and ADVIA Centaur/ ADVIA centaur XP/ ADVIA Centaur XPT systems interchangeably when interpreting serial testing in patient samples. This communication does not impact customers who use TnI-Ultrat results solely from either the ADVIA Centaur/ ADVIA Centaur XP/ ADVIA Centaur XPT system or the ADVIA Centaur CP system.