Health Products

AMYL2 and LDHI2 for cobas C701 Analyzer (2015-05-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
B. LDHI2  for cobas C701 Analyzer5169330190
All lots.
Manufacturer
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
A. AMYL2 for cobas C701 Analyzer5167027190

Introduction

Recalled Products

A. AMYL2 for cobas C701 Analyzer
B. LDHI2  for cobas C701 Analyzer

Reason

Potential to obtain falsely low results without a flag in samples with high turbidity and high kinetic activity. This issue affects only LDHI2 and AMYL2 assays.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime