Health Products

CADD-Solis Ambulatory Infusion Pumps (2015-06-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B. CADD-SOLIS Ambulatory Infusion Pumps
Manufacturer
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES
21-2111-0200-02                       
More than 10 numbers, contact manufacturer.A. CADD-SOLIS VIP Ambulatory Infusion Pump
Manufacturer
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES
21-2120-0102-02                       

Introduction

Recalled Products

A. CADD-SOLIS VIP Ambulatory Infusion Pump
B. CADD-SOLIS Ambulatory Infusion Pumps

Reason

Smiths Medical reports the intermittent occurrence of binding of the locking assembly on some CADD-Solis pumps. Binding on the cassette/keypad lock can occur after latching the CADD-Solis medication cassette reservoir or administration set disposable to the pump. When binding occurs, it can prevent the key from fully rotating the cassette/keypad lock to the locked (engaged) position. To start an infusion with pump models 2111 and 2112, and model 2120, when in PCA mode, the lock must be fully engaged. This issue is limited to specific serial numbers only (1061043 through 1067598).
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime