Dimension Vista System (2015-05-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 Gbc Drive, Mailstop 514
Newark, Delaware
19714-6101
UNITED STATES
B. Dimension Vista System - Dimension Vista1500Not applicable
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 Gbc Drive, Mailstop 514
Newark, Delaware
19714-6101
UNITED STATES
C. Dimension Vista System - Dimension Vista 5001000034648
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 Gbc Drive, Mailstop 514
Newark, Delaware
19714-6101
UNITED STATES
A. Dimension Vista System - Dimension Vista 1500 Analyzer1000033793

Introduction

Recalled Products
  1. Dimension Vista System - Dimension Vista 1500 Analyzer
  2. Dimension Vista System - Dimension Vista1500
  3. Dimension Vista System - Dimension Vista 500

Reason

Issue # 1: Manufacturer received complaints for samples not processing without notification on the Dimension Vista 500 or Dimension Vista 1500 using software versions 3.6.1 in most cases, user action of pause / cancel pause is effective in restoring lane operation and the samples will resume processing. However in a few cases it was necessary to shutdown the instrument software and restart to recover. In all cases the affected samples are resubmitted and run successfully.

Issue # 2: Manufacturer received complaints for unexpected less than lower assay range (< l) or results that were flagged with assay errors on Dimension Vista 1500 systems using software versions 3.6.1. Result data and the physical evidence from flexes suggested that a reagent server lost synchronization and the software was unaware of the problem.