A. Dimension Vista System - Dimension Vista 1500 Analyzer
1000033793
Introduction
Recalled Products
Dimension Vista System - Dimension Vista 1500 Analyzer
Dimension Vista System - Dimension Vista1500
Dimension Vista System - Dimension Vista 500
Reason
Issue # 1: Manufacturer received complaints for samples not processing without notification on the Dimension Vista 500 or Dimension Vista 1500 using software versions 3.6.1 in most cases, user action of pause / cancel pause is effective in restoring lane operation and the samples will resume processing. However in a few cases it was necessary to shutdown the instrument software and restart to recover. In all cases the affected samples are resubmitted and run successfully.
Issue # 2: Manufacturer received complaints for unexpected less than lower assay range (< l) or results that were flagged with assay errors on Dimension Vista 1500 systems using software versions 3.6.1. Result data and the physical evidence from flexes suggested that a reagent server lost synchronization and the software was unaware of the problem.