Health Products

REMEL PATHODX STREP GROUP B LATEX (2015-05-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
LOT 1463129
Manufacturer
Oxoid Limited
Wade Road
Basingstoke, Hampshire
RG24 8PW
UNITED KINGDOM
A.REMEL PATHODX STREP GROUP B LATEXR62025

Introduction

Recalled Product

A.REMEL PATHODX STREP GROUP B LATEX
 

Reason

A technical investigation has confirmed that PathoDX Strep B Latex may produce weak or slow reactions that may result in a failure to identify b-haemolytic streptococci group b isolates potentially leading to a delay in result reporting and/or a delay to delivering appropriate therapy.