AUTOMATE 2500 SYSTEM (2014-10-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
2205A.AUTOMATE 2500 SYSTEM
Manufacturer
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
ODL25250

Introduction

Recalled Product

A.AUTOMATE 2500 SYSTEM
 

Reason

An IR sensor module failure can occur and can be identified by one of the following: volume detection error message ErrID: 5694, timeout error message ErrID: 5405, system initialization failure of the conveyor belt module. Immediately prior to the failure there is a remote possibility that the IR sensor module may not properly detect the sample volume in the tube. The IR sensor module is part of the Tube Inspection Unit (TIU) and is responsible for sample volume detection on AutoMate systems. Failure to properly detect the sample volume in the tube may cause an increased number of primary and secondary sample tubes to be transferred to the error workplace on the AutoMate system. As a result of the failure, there may be unexpected downtime of the AutoMate system.