B. CONTOUR NEXT CONTROL SOLUTION (HIGH)
7324
All lots.
Manufacturer
Bayer Healthcare LLC Subsidiary of Bayer Corporation
P.O. Box 2004, 430 S. Bieger Street
Mishawaka
Indiana
UNITED STATES
C. CONTOUR NEXT CONTROL SOLUTION (LOW)
7323
All lots.
Manufacturer
Bayer Healthcare LLC Subsidiary of Bayer Corporation
P.O. Box 2004, 430 S. Bieger Street
Mishawaka
Indiana
UNITED STATES
A. CONTOUR NEXT CONTROL SOLUTION (NORMAL)
7325
Introduction
Recalled Products
CONTOUR NEXT CONTROL SOLUTION (NORMAL)
CONTOUR NEXT CONTROL SOLUTION (HIGH)
CONTOUR NEXT CONTROL SOLUTION (LOW)
Reason
Control result not marked as a control test: Bayer Healthcare has received user reports on CONTOUR NEXT control test results that are higher than the control range published on the packaging of contour next test strips and not recognized as control tests by the meter. All lots of CONTOUR NEXT control solution (levels low, normal and high) are affected.
Control result outside of control range: Another issue may affect any control solution lots when used with of CONTOUR NEXT test strip lots with expiration dates through November 2016. Due to unforeseen material variation, lots of contour next control solution may give control test results that are outside of the assigned control range as published on the packaging of CONTOUR NEXT test strips. These control results are properly marked as control tests by the meter therefore this issue does not impact insulin pumps.