Health Products

COBAS 8000 DATA MANAGER (2015-05-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N/A
Manufacturer
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
A. COBAS 8000 DATA MANAGER6888160001

Introduction

Recalled Product

A. COBAS 8000 DATA MANAGER

Reason

Potential issue with the Cobas 8000 data manager system (all versions) where the option "automatic patient masking when QC error" option does work correctly. This functionality does not work if the results for the last measured control level for this test is valid. As a consequence, the results for the measured samples would not be released and forwarded to host due to the QC error.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime