Health Products

MOBILEDIAGNOST WDR (2015-06-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
ALLA. MOBILEDIAGNOST WDR
Manufacturer
Philips Medical Systems Dmc Gmbh
Rontgenstrabe 24
Hamburg
22335
GERMANY
9890 010 8952X

Introduction

Recalled Product

A. MOBILEDIAGNOST WDR

Reason

Philips has become aware that when the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle. Proper behavior is re-enabled when the system is rebooted.