ATTUNE Tibial Articulation Surface Instruments (2015-06-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
>100 NUMBERS, CONTACT MFR
Manufacturer
Depuy (Ireland)
Loughbeg
Ringaskiddy
IRELAND
D. ATTUNE RP CR ARTICULATION SURFACE2545-00-563
>100 NUMBERS, CONTACT MFR
Manufacturer
Depuy (Ireland)
Loughbeg
Ringaskiddy
IRELAND
E. ATTUNE CONV FB PS ARTICULATION SURFACE2545-00-504
>100 NUMBERS, CONTACT MFR
Manufacturer
Depuy (Ireland)
Loughbeg
Ringaskiddy
IRELAND
B. ATTUNE CONV RP PS ARTICULATION SURFACE254500993
>100 NUMBERS, CONTACT MFR
Manufacturer
Depuy (Ireland)
Loughbeg
Ringaskiddy
IRELAND
C. ATTUNE CONV RP PS TB TRIAL254500996
>100 NUMBERS, CONTACT MFR
Manufacturer
Depuy (Ireland)
Loughbeg
Ringaskiddy
IRELAND
A.ATTUNE CONV FB PS TB TRIAL254500973

Introduction

Recalled Product
  1. ATTUNE CONV FB PS TB TRIAL
  2. ATTUNE CONV RP PS ARTICULATION SURFACE
  3. ATTUNE CONV RP PS TB TRIAL
  4. ATTUNE RP CR ARTICULATION SURFACE
  5. ATTUNE CONV FB PS ARTICULATION SURFACE

Reason

DePuy is issuing a device correction for all lots of ATTUNE Tibial Articulation Surface Instruments used in trialing with the ATTUNE knee system. The device correction is being issued due to the potential for the balseal, which is a small wire spring coil located on the post feature of the Articulation Surface Instruments, to become damaged and disassociate (come off). The balseal is constructed from implant grade stainless steel. If the balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.