Health Products

WIZARD2 GAMMA COUNTER (2015-06-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
>10 NUMBERS, CONTACT MFGA. WIZARD2 GAMMA COUNTER 
Manufacturer
Perkinelmer Singapore Pte Ltd.
BLK 28 Ayer Rajah Crescent, #04-01/08
13959
SINGAPORE
2470-0010

Introduction

Recalled Product

A. WIZARD2 GAMMA COUNTER

Reason

Manufacturer has become aware through a customer complaint that the content of the second of the two #023 barcode id labels provided on each page of the barcode ID label binder are incorrect. the second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. in the event measurement protocols have been programmed for both ID #023 and #024, and the WIZARD2 Gamma Counter scans the affected barcode label, the protocol associated with barcode ID #024 is executed. If the error is undetected, the WIZARD2 Gamma Counter may produce erroneous results.