Synthes Canada, Ltd. is initiating a voluntary medical device labelling recall for Synthes Reamer/Irrigator /Aspirator (RIA) system technique guide (J4352). Precautionary statements are being added to the "Reamer /Irrigator/ Aspirator (RIA) surgical technique guide". It was reported that the ria drive shaft, tube assembly, and reamer head have the potential to break when incorrectly assembled or used improperly as follows: 1. failure to fully engage the reamer to shaft at assembly prior to surgery. Incorrect assembly creates a reduced reamer/shaft contact surface area, in which excess rotational forces could cause breakage. 2. Use of RIA drive shafts after the flats has become worn and rounded. 3. Use of any excess force beyond design limit 4. Use of with an incompatible power tool.