A. VERSATREK AUTOMATED MICROBIAL DETECTION SYSTEM (528 MODEL)
Manufacturer
Remel Inc.
12076 Santa Fe Drive
Lenexa
66215
Kansas
UNITED STATES
6528-22
Introduction
Recalled Product
A.VERSATREK AUTOMATED MICROBIAL DETECTION SYSTEM (528 MODEL)
Reason
An internal investigation has shown that the use of the VersaTREK Automated Microbial Detection System might result in false positive reports. To date three (3) customer complaints have been received, all in conjunction with VersaTREK Myco bottle usage. Review of instrument databases suggests that on rare occasions the barometric pressure readings may be inconsistent. This event is caused by a protective sticker covering the sensor. If this sticker forms a complete seal it traps air under the sticker, resulting in incorrect atmospheric pressure readings. This in turn leads to reports of false positives.