Health Products

VERSATREK AUTOMATED MICROBIAL DETECTION SYSTEM (528 MODEL) (2015-05-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
2298X0714A. VERSATREK AUTOMATED MICROBIAL DETECTION SYSTEM (528 MODEL)
Manufacturer
Remel Inc.
12076 Santa Fe Drive
Lenexa
66215
Kansas
UNITED STATES
6528-22

Introduction

Recalled Product

A.VERSATREK AUTOMATED MICROBIAL DETECTION SYSTEM (528 MODEL)

Reason

An internal investigation has shown that the use of the VersaTREK Automated Microbial Detection System might result in false positive reports. To date three (3) customer complaints have been received, all in conjunction with VersaTREK Myco bottle usage. Review of instrument databases suggests that on rare occasions the barometric pressure readings may be inconsistent. This event is caused by a protective sticker covering the sensor. If this sticker forms a complete seal it traps air under the sticker, resulting in incorrect atmospheric pressure readings. This in turn leads to reports of false positives.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime